wpe1.gif (5818 bytes)

wpe2.gif (40141 bytes) The Jerome 431-X Mercury Vapor Analyzer uses a gold film sensor for the detection and accurate measurement of toxic
mercury vapor in the air.

A portable hand-held unit, the Jerome 431-X can easily be carried to locations of mercury concerns. Simple, push-button
operation allows users to measure mercury levels in just seconds. The range of the 431-X is from 0.000 to 0.999 mg/m3 Hg.
The gold film sensor is inherently stable and selective to mercury, eliminating interference common to ultraviolet analyzers, such
as water vapor and hydrocarbons.

The 431-X includes features not available in older Jerome models. When attached to either a data logger or computer, the
analyzer automatically regenerates the sensor when it becomes saturated and then resumes sampling. An improved film
regeneration circuit makes the sensor last even longer.
Features Portable, rugged and easy to operate
13 second response
Automated sample cycle
Inherently stable gold film sensor
Operates up to six hours on fully charged nickel-cadmium batter
Digital display in either mg/m3 or ng
One year limited parts and labor warranty
Applications Mercury surveys
Spill response
Hazardous waste sites
Mercury exclusion tests
Specifications Accuracy: +/- 5% at 0.100 mg/m3 Hg
Sensitivity: 0.003 mg/m3 Hg
Range 0.000 to 0.999 mg/m3 Hg
Precision: 5% relative standard deviation at 0.100 mg/m3 Hg
Environmental Operating Range: 0-40 degrees C. non-condensing, non-explosive
Response Time: Sample mode - 13 seconds; Survey mode - 4 seconds
Flow Rate: 750cc/min.
Weight: 7 pounds
Dimensions: 6"W x 13"L x 4"H
Case Design: Enform
Power Requirements: 115VAC or 230VAC, 115 watts maximum